Key result
Replacement of the broken pacemaker electrode successfully restored pacing and resolved the patient's recurrent syncope.
Why the study?
In patients with a narrow space between the clavicle and first rib, friction from upper limb activity can cause pacemaker electrode wire breakage, resulting in poor sensing and pacing.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
In patients with pacemakers implanted via the subclavian approach who present with recurrent syncope, electrode rupture due to friction should be considered and treated with electrode replacement.
BACKGROUND: Currently, the implantation of permanent cardiac pacemakers entails mostly subclavian vein puncture, which is relatively simpler and easier to master. However, due to individual differences, some patients carry a narrow space between the clavicle and the first rib. If the range of activity of the upper limb is increased, the friction between the electrode wire and the bone gap leads to the breakage of the electrode wire, which is manifested by poor pacemaker perception and pacing. CASE SUMMARY: A 68-year-old woman underwent permanent pacemaker implantation in our hospital because of third-degree atrioventricular block 6 years ago. At that time, the patient was recommended to have a dual-chamber permanent pacemaker implantation, and finally chose a single-chamber permanent pacemaker because she could not afford the cost. The patient has repeatedly lost consciousness for no obvious reason in the past 3 d, and went to our hospital for treatment. The chest X-ray showed that the pacemaker electrode was broken. After the patient was given a pacemaker electrode replacement, the patient did not continue to lose consciousness. CONCLUSION: Because the electrodes implanted in the subclavian approach are close to the clavicle and the first rib, the pacemaker electrodes may wear out. If the patient loses consciousness again after the pacemaker is implanted, we should consider whether there is a pacemaker. The possibility of electrode breakage, and timely help the patient to replace the new pacemaker electrodes.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2022) conducted a case report in Recurrent syncope due to pacemaker electrode rupture (n=1). Pacemaker electrode replacement was evaluated on Resolution of syncope. Replacement of the broken pacemaker electrode successfully restored pacing and resolved the patient's recurrent syncope.
Synapse has enriched 10 related papers. Consider them for comparative context: