Key result
Emergency implantation of a new pacemaker on the contralateral side successfully treated positional convulsant syncope caused by a fractured right ventricular lead in a 73-year-old man.
Case Report (n=1)
Highlights pacemaker lead fracture as a potential cause of positional convulsant syncope in patients with older pacemakers.
Answers are grounded in this paper and cite their sources.
Consider lead fracture in positional syncope with pacemakers; single case leaves open generalizability and optimal strategies.
In spite of the advances made in the technology of pacemakers which resulted in a decrease in the incidence of pacemaker lead fracture, the latter remains a potential complication of implanted pacemakers manufactured in the early days. In this report, we present a case of fracture of the unipolar electrode diagnosed by an emergency physician in a patient on a pacemaker for 10 years who presented to the emergency department with positional convulsant syncopes.
No takes yet. Share an insight, caveat, or question.
Lassoued et al. (2014) conducted a case report in Positional convulsant syncope due to pacemaker lead fracture (n=1). Emergency implantation of a new pacemaker was evaluated on Clinical outcome. Emergency implantation of a new pacemaker on the contralateral side successfully treated positional convulsant syncope caused by a fractured right ventricular lead in a 73-year-old man.