Key result
The Cingular bovine pericardial aortic valve demonstrated 5-year freedom from all-cause mortality of 91.2%, structural valve deterioration of 100%, and all-cause reintervention of 99.3%.
Why the study?
Although the short-term performance of the Cingular bovine pericardial aortic valve had been demonstrated, its 5-year safety and haemodynamic outcomes remained to be evaluated.
What are the 5-year safety and haemodynamic outcomes of surgical aortic valve replacement with the Cingular bovine pericardial aortic valve?
Cohort (n=148)
Yes
What are the 5-year safety and haemodynamic outcomes of surgical aortic valve replacement with the Cingular bovine pericardial aortic valve?
Answers are grounded in this paper and cite their sources.
The Cingular bovine pericardial aortic valve demonstrated excellent 5-year safety and haemodynamic performance, with 100% freedom from structural valve deterioration.
OBJECTIVES: The short-term performance of the Cingular bovine pericardial aortic valve was proven. This study evaluated its 5-year safety and haemodynamic outcomes. METHODS: It enrolled 148 patients who underwent surgical aortic valve replacement with the Cingular bovine pericardial aortic valve between March 2016 and October 2017 in 5 clinical centres in China. Safety and haemodynamic outcomes were followed up to 5 years. The incidence of all-cause mortality, structural valve deterioration and reintervention was estimated by Kaplan-Meier analysis. RESULTS: The mean age of patients was 67.7 [standard deviation (SD) 5.1] years, and 36.5% of patients were female. The mean follow-up was 5.3 (SD 1.2) years. Five-year freedom from all-cause mortality, structural valve deterioration and all-cause reintervention were 91.2%, 100% and 99.3%, respectively. At 5 years, the mean gradient and effective orifice area of all sizes combined were 14.0 (SD 5.5) mmHg and 1.9 (SD 0.3) cm2, respectively. For 19- and 21-mm sizes of aortic prostheses, the mean gradients and effective orifice area at 5 years were 17.5 (SD 7.0) mmHg and 1.6 (SD 0.2) cm2 and 13.7 (SD 6.7) mmHg and 1.8 (SD 0.3) cm2, respectively. The incidence of moderate or severe patient-prosthesis mismatch was 4.1% and 0.0% patients at 5 years, respectively. CONCLUSIONS: The 5-year safety and haemodynamic outcomes of Cingular bovine pericardial aortic valve are encouraging. Longer-term follow-up is warranted to assess its true durability.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2024) conducted a cohort in Aortic valve disease requiring surgical replacement (n=148). Cingular bovine pericardial aortic valve was evaluated on 5-year freedom from all-cause mortality. The Cingular bovine pericardial aortic valve demonstrated 5-year freedom from all-cause mortality of 91.2%, structural valve deterioration of 100%, and all-cause reintervention of 99.3%.
Synapse has enriched 10 related papers. Consider them for comparative context: