Key result
Right ventricular septal pacing and apical pacing yielded similar maximum stroke volumes at optimal sensed and paced atrioventricular delays.
Why the study?
Little is known regarding the appropriate atrioventricular delay between right ventricular septal and right ventricular apical pacing for optimal hemodynamic benefit.
Are there differences in the optimal atrioventricular delays required to maximize stroke volume between right ventricular septal and apical pacing in patients with complete AV block and normal LV function?
Observational (n=50)
Are there differences in the optimal atrioventricular delays required to maximize stroke volume between right ventricular septal and apical pacing in patients with complete AV block and normal LV function?
Answers are grounded in this paper and cite their sources.
The optimal atrial sensed and paced AV delays for maximizing hemodynamics are similar between right ventricular septal and apical pacing in patients with normal left ventricular function.
INTRODUCTION: There is a surge of interest in alternate site pacing to prevent pacing-induced left ventricular dysfunction. However, little is known regarding the appropriate atrioventricular (AV) delay between right ventricular (RV) septal and RV apical pacing for optimal hemodynamic benefit. OBJECTIVES: To determine the programmed values of atrial sensed and atrial paced AV delays in basal RV septal and apical RV pacing that results in the maximum delivered stroke volume (SV). METHODS: We calculated the Doppler-derived SV at various sensed and paced AV delays in 50 patients with complete AV block implanted with a dual-chamber pacemaker (group A: 25 RV apical pacing; group B: 25 RV septal pacing). The hemodynamic difference in terms of the SV between sensed and paced AV delay, corresponding to the site of RV pacing was then compared for statistical significance. RESULTS: In group A, maximal SV was derived at a sensed AV delay of 123.2 ± 11 ms and paced AV delay of 129.2 ± 10 ms, and in group B, at a sensed AV delay of 123.6 ± 8 ms and paced AV delay of and 132.8 ± 7 ms. At these intervals, there was no difference in the SV between septal and apical RV pacing (P = .28 and .22, respectively). CONCLUSION: The atrial sensed and atrial paced AV delays for septal and apical RV pacing for optimal hemodynamics are similar. For optimal hemodynamics, the atrial paced AV delay is longer than the atrial sensed AV delay.
No takes yet. Share an insight, caveat, or question.
Shenthar et al. (2019) conducted an observational in Complete AV block (n=50). Right ventricular septal pacing vs. Right ventricular apical pacing was evaluated on Maximum delivered stroke volume (SV) (p=0.28 and 0.22). Right ventricular septal pacing and apical pacing yielded similar maximum stroke volumes at optimal sensed and paced atrioventricular delays.
Synapse has enriched 10 related papers. Consider them for comparative context: