Key result
Guidewire insertion during percutaneous coronary intervention in a patient with an ICD caused noise leading to an inappropriate shock.
Case Report (n=1)
Physicians performing percutaneous coronary interventions in patients with ICDs or pacemakers must be aware of the risk of inappropriate shocks or pacing inhibition caused by procedural noise.
Answers are grounded in this paper and cite their sources.
Case report of ICD shock from PCI guidewire noise; hypothesis-generating for procedural precautions in device patients.
Coronary disease is a common condition in patients affected by heart failure with severely reduced ejection fraction (HFrEF). This condition represents an indication for implantable cardioverter defibrillator (ICD) in order to reduce the risk of sudden death related to arrhythmias. Nevertheless, inappropriate shocks are associated with worse quality of life, hospitalization, and death. We present the case of an inappropriate shock related to percutaneous coronary intervention during the insertion and advancement of the guidewire into the left anterior descending artery (LAD) in a patient with an ICD. Physicians' awareness about the clinical implication of noise arising during a coronary procedure is very important in patients with an ICD or pacemaker, to avoid inappropriate shock or pacing inhibition and to raise the possibility of lead implantation in or helix protrusion into the coronary lumen.
No takes yet. Share an insight, caveat, or question.
Stolfo et al. (2019) conducted a case report in Heart failure with severely reduced ejection fraction (HFrEF) and an ICD (n=1). Percutaneous coronary intervention (guidewire insertion) was evaluated on Inappropriate shock. Guidewire insertion during percutaneous coronary intervention in a patient with an ICD caused noise leading to an inappropriate shock.
Synapse has enriched 10 related papers. Consider them for comparative context: