Key result
Biventricular sensing and pacing demonstrated comparable median induced VF detection times and first-attempt VT termination rates compared to RV pacing alone.
Why the study?
Does an ICD capable of biventricular pacing and sensing improve detection times of induced VF and success rates for terminating induced VT compared to an ICD using an RV lead alone in heart failure patients with VT/VF and QRS >= 120 ms?
Cohort (n=219)
Does an ICD capable of biventricular pacing and sensing improve detection times of induced VF and success rates for terminating induced VT compared to an ICD using an RV lead alone in heart failure patients with VT/VF and QRS >= 120 ms?
Biventricular sensing and pacing in ICDs is as effective as conventional right ventricular sensing and pacing for detecting VF and terminating VT.
Answers are grounded in this paper and cite their sources.
Comparable BiV sensing/pacing performance in induced testing supports CRT-D use; hypothesis-generating and should not yet change practice.
While much is known concerning the hemodynamic effects of biventricular (BV) pacing, little has been reported concerning the efficacy of BV sensing and pacing in the detection and treatment of ventricular tachyarrhythmias. Two hundred nineteen heart failure (HF) patients with VT or VF and a QRS > or = 120 ms during sinus rhythm received an ICD capable of BV pacing and sensing. Detection times of induced VF and success rates for terminating induced VT were measured. The ICD system used a left ventricular epicardial lead implanted via thoracotomy (52 patients) or a specially designed percutaneous, over-the-wire lead inserted in the coronary venous system. VF detection times and VT termination rates by antitachycardia pacing (ATP) were compared with those measured in a population of recipients of ICD using a RV lead alone. Median induced VF detection times were comparable (2.0-s BV vs 1.8-s RV). Termination of induced VT on the first attempt was comparable with BV pacing (87.4%) versus RV pacing (89.6%). The time to detect induced VF was not different with ICDs using BV sensing versus conventional ICDs using RV sensing alone. Similarly, the rates of successful termination of induced VT by ATP with BV or RV pacing were comparable.
No takes yet. Share an insight, caveat, or question.
Bocchiardo et al. (2000) conducted a cohort in Heart failure with VT or VF (n=219). Biventricular (BV) sensing and pacing ICD vs. ICD using a right ventricular (RV) lead alone was evaluated on Termination of induced VT on the first attempt. Biventricular sensing and pacing demonstrated comparable median induced VF detection times and first-attempt VT termination rates compared to RV pacing alone.