Key result
Ten years after surgical aortic valve replacement with the Perimount Magna Ease prosthesis, the cumulative incidence of any structural valve deterioration was 25% and severe deterioration was 14%.
Why the study?
To evaluate the long-term clinical outcome, hemodynamic performance, and the incidence and course of structural valve deterioration of the Carpentier-Edwards Perimount Magna Ease bioprosthesis used for SAVR.
Cohort (n=58)
No
The Perimount Magna Ease bioprosthesis demonstrates excellent 10-year hemodynamic performance, with severe structural valve deterioration occurring in 14% of patients at 10 years.
Background: The Carpentier-Edwards Perimount Magna Ease prosthesis (PME) represents the latest generation of stented bioprostheses used for surgical aortic valve replacement (SAVR). The aim of our study was to evaluate the long-term clinical outcome and hemodynamic performance of the prosthesis with a focus on the incidence and course of structural valve deterioration (SVD) by serial echocardiographic examinations.Methods: SAVR with the PME was performed in 58 consecutive patients between 2007 and 2008. Transthoracic echocardiography was performed preoperatively, at discharge and annually during a 10-year follow-up at the German Heart Center Munich.Results: Mean age at surgery was 62±14 years. At discharge (n=57), the overall mean pressure gradient (MPG) and effective orifice area (EOA) were 15.8±4.1 mmHg and 1.8±0.4 cm2, respectively. Moderate patient-prosthesis mismatch (PPM) was present in 18 patients (32%) and severe PPM in 6 patients (11%) at discharge. Ten years following SAVR (n=33), the overall MPG was 16.6±7.3 mmHg and EOA was 1.3±0.4 cm2.Thirty-day and late mortality was 2% (n=1) and 21% (n=12), respectively. Survival at 1, 5, and 10 years was 94.7%±3.3%, 91.1%±4.1%, and 77.3%±5.9%, respectively. Freedom from reoperation at 10 years was 88.8%±4.7%. Ten years after PME implantation the cumulative incidence of any SVD, severe SVD, and bioprosthetic valve failure (BVF) was 25%±6%, 14%±5%, and 16%±5%, respectively.Conclusions: The PME shows an excellent hemodynamic performance over the course of 10 years with development of clinically relevant SVD as late as 6 years post implant, and a 10-year incidence of severe SVD of 14%.
No takes yet. Share an insight, caveat, or question.
Mayr et al. (2021) conducted a cohort in Surgical aortic valve replacement (n=58). Carpentier-Edwards Perimount Magna Ease prosthesis was evaluated on Cumulative incidence of any structural valve deterioration (SVD) at 10 years. Ten years after surgical aortic valve replacement with the Perimount Magna Ease prosthesis, the cumulative incidence of any structural valve deterioration was 25% and severe deterioration was 14%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: