Key result
SyncAV dynamic AV-delay optimization causes expected transient QRS changes mimicking loss of biventricular pacing.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that mimic loss of biventricular pacing, and recognizing this expected behavior prevents unnecessary device evaluation.
CASE SUMMARY: A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction. TAKE-HOME MESSAGE: Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.
No takes yet. Share an insight, caveat, or question.
Weber et al. (2026) conducted a case report in nonischemic cardiomyopathy and left bundle branch block (n=1). Dynamic atrioventricular-delay optimization algorithms (SyncAV) was evaluated. Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that mimic loss of biventricular pacing but represent expected device behavior.