Key result
Noninvasive CRT programming successfully suppressed right ventricular lead-induced ventricular arrhythmia without the need for catheter ablation.
Why the study?
Can noninvasive CRT-D programming suppress right ventricular lead-induced ventricular arrhythmia in a patient with heart failure?
Case Report (n=1)
Can noninvasive CRT-D programming suppress right ventricular lead-induced ventricular arrhythmia in a patient with heart failure?
Noninvasive CRT programming can effectively suppress rare right ventricular lead-induced ventricular arrhythmias without requiring catheter ablation.
May allow noninvasive management of RV lead-induced VA in HF; leaves open need for prospective validation.
BACKGROUND: A 53-year-old male with heart failure secondary to anterior wall myocardial infarction treated with cardiac resynchronization-defibrillator (CRT-D) device presented with ventricular arrhythmia: repetitive incessant slow ventricular tachycardias (VT) below the CRT-D detection zone, accelerated ventricular rhythm, and numerous premature ventricular ectopic beats (ExV), resulting in loss of biventricular pacing. METHODS AND RESULTS: Nonsustained monomorphic VT (nsVT) and ExV were observed in an electrocardiogram under biventricular stimulation. During noninvasive CRT-D programming, ventricular bigeminy reproducibly recurred only at right ventricular (RV) pacing and its morphology was almost identical to the stimulated beats. The left ventricular (LV) pacing failed to induce ventricular ectopy or tachycardia. CONCLUSIONS: This unusual case shows a rare phenomenon of late proarrhythmic effect due to the RV lead pacing-a new finding reported only in a few publications. Here we present our approach to CRT programming that suppressed the clinical arrhythmia without the need of catheter ablation and achieving the high biventricular pacing capture rate along with optimal hemodynamic CRT-D performance.
No takes yet. Share an insight, caveat, or question.
Miśkowiec et al. (2019) conducted a case report in Heart failure with CRT-D presenting with ventricular arrhythmia (n=1). Noninvasive CRT programming was evaluated on Suppression of clinical arrhythmia. Noninvasive CRT programming successfully suppressed right ventricular lead-induced ventricular arrhythmia without the need for catheter ablation.