Key result
In an in vitro hydraulic circuit, adjusting upstream and downstream aeration volumes to 360 mL and 180 mL, respectively, achieved an emulational flow that fit realistic human circulatory wave forms.
This in vitro study demonstrates that adjusting upstream and downstream compliance in a hydraulic circuit can accurately emulate realistic human blood flow waveforms for hemodynamic research.
Answers are grounded in this paper and cite their sources.
May refine in vitro hemodynamic models for device testing; leaves open in vivo validation.
Generating an artificial blood flow in a circle system is an important step in hemodynamic research; thus, the influence of circle system components on the pulsatile flow wave forms should be investigated. In this study, a circle system was built, in which two solenoid valves were controlled by a timer to transform a constant flow into a pulsatile one, and two customized compliances with five different aeration volumes were set up upstream and downstream of the test chamber, achieving twenty-five different wave forms. Then, the influence of the compliance settings on the flow rate wave form was investigated. From the experimental data, it can be concluded that the absolute value of the maximum value and the minimum value of the wave forms increases along the aeration of the downstream compliance but decreases along the aeration of the upstream compliance. For the second maximum value and the offset between the maximum value and the minimum value, remarkable differences were obtained between runs with aeration of zero compliance and nonzero compliance. Finally, an emulational flow was achieved with the up- and downstream aeration volume equaling 360 mL and 180 mL, respectively, which fit the realistic wave form well.
No takes yet. Share an insight, caveat, or question.
He et al. (2014) studied this question. Compliance settings (aeration volumes) vs. different aeration volumes (zero vs nonzero compliance) was evaluated on Flow rate wave forms. In an in vitro hydraulic circuit, adjusting upstream and downstream aeration volumes to 360 mL and 180 mL, respectively, achieved an emulational flow that fit realistic human circulatory wave forms.
Synapse has enriched 10 related papers. Consider them for comparative context: