Key result
A standardized pressure optimization protocol for stent deployment required a mean inflation time of 104 ± 41 seconds to achieve pressure stability, with stent length predicting prolonged inflation.
Why the study?
Does a standardized pressure optimization protocol determine the required duration of stent inflation in patients undergoing stent implantation?
Observational (n=435)
Does a standardized pressure optimization protocol determine the required duration of stent inflation in patients undergoing stent implantation?
Achieving stable inflation pressure during stent deployment requires an average of over 100 seconds, suggesting that the common practice of rapid inflation/deflation may be insufficient for full stent expansion.
Answers are grounded in this paper and cite their sources.
POP may optimize stent expansion versus rapid inflation; leaves open whether it improves outcomes in randomized trials.
BACKGROUND: High-pressure inflation is the universal standard for stent deployment but a specific protocol for its use is lacking. We developed a standardized "pressure optimization protocol" (POP) using time to inflation pressure stability as an endpoint for determining the required duration of stent inflation. OBJECTIVES: The primary study purpose was to determine the stent inflation time (IT) in a large patient cohort using the standardized inflation protocol, to correlate various patient and lesion characteristics with IT, and ascertain in an in vitro study the time for pressure accommodation within an inflation system. METHODS: Six hundred fifteen stent implants in 435 patients were studied. Multivariate analysis was performed to determine predictors of longer ITs. In an in vitro study, various stents and balloons were inflated in air to determine the pressure accommodation time of the inflation system. RESULTS: The mean stent IT was 104 ± 41 sec (range 30-380 sec). Stent length was the only predictor of prolonged stent inflation. The "accommodation time" in vitro of the stent inflation system itself was 33 ± 24 sec. The protocol was safe requiring premature inflation termination in <3% of stent implants. No serious adverse events occurred. CONCLUSIONS: Achieving stable inflation pressure requires on average over 100 sec and may require several minutes in individual cases. Stent length increases IT. These results suggest that the widespread practice of rapid inflation/deflation may not be sufficient to fully expand the stent and that the use of a pressure stability protocol will allow for safe, predictable, and more complete stent deployment.
No takes yet. Share an insight, caveat, or question.
Vallurupalli et al. (2015) conducted an observational in Stent implantation (n=435). Pressure optimization protocol (POP) was evaluated on stent inflation time (IT). A standardized pressure optimization protocol for stent deployment required a mean inflation time of 104 ± 41 seconds to achieve pressure stability, with stent length predicting prolonged inflation.