Answers are grounded in this paper and cite their sources.
The apparatus consists of a small, electrically heated thermocouple bent into a loop at its extremity so as to hold the specimen, in the form of a bead, in contact with the thermojunction. It is based on the apparatus described by Ordway* but has new and simpler electrical arrangements. For convenience, the bead is normally observed through a 25 mm microscope objective, though a 16 mm objective (5 mm working distance) can be used without any special precautions. The heating current is taken from any convenient 50 c/s supply and a vibrating switch driven from the same supply permits the heating current to flow only during alternate half-cycles, the measuring circuit being employed during the intermediate half-cycles. The power dissipated by the heater at maximum temperature (1750° C) is only about 6 W. The apparatus gives extremely rapid results in comparison with normal techniques when put to such uses as melting-point determinations, studies of glass devitrification, growing single crystals for X-ray structure determinations, and delineation of phase equilibria; it should find widespread application in the fields of glass and ceramic technology.
No takes yet. Share an insight, caveat, or question.
Jane Welch (1954) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: