Key result
Bedside embolectomy catheter atrioseptostomy successfully enlarges restrictive ASD and raises oxygen saturation in a preterm infant.
Why the study?
Interventional treatment of restrictive atrial septal defects in complex heart disease is state-of-the-art therapy, but dedicated balloons are lacking after several products were withdrawn from the market.
Can an off-label Edwards Fogarty Fortis Arterial Embolectomy Catheter safely and effectively achieve bedside echocardiography-guided balloon-atrioseptostomy in a low-birth-weight preterm infant with transposition of the great arteries and a restrictive atrial septal defect?
Case Report (n=1)
No
Can an off-label Edwards Fogarty Fortis Arterial Embolectomy Catheter safely and effectively achieve bedside echocardiography-guided balloon-atrioseptostomy in a low-birth-weight preterm infant with transposition of the great arteries and a restrictive atrial septal defect?
Demonstrates the feasibility of off-label bedside echo-guided balloon-atrioseptostomy using a shapeable arterial embolectomy catheter in a 1.2 kg preterm infant when dedicated BAS catheters are unavailable or anatomically unsuitable.
Answers are grounded in this paper and cite their sources.
May offer a rescue option when standard catheters are unavailable; leaves open prospective safety evaluation before wider use.
Interventional treatment of restrictive atrial septal defects in complex heart disease is considered state-of-the-art therapy up to date. Nevertheless, dedicated balloons are lacking so far, as several products have been withdrawn from the market. We report on off-label use of a balloon embolectomy catheter used successfully in a preterm patient and discuss whether this device might be used in other patients as well as it seems to be promising due to its shape and versatility.
No takes yet. Share an insight, caveat, or question.
Pozza et al. (2023) conducted a case report in Transposition of the great arteries with ventricular septal defect and restrictive atrial septal defect (n=1). Bedside echocardiography-guided balloon-atrioseptostomy using an Edwards Fogarty Fortis Arterial Embolectomy Catheter was evaluated on Atrial septal defect diameter and systemic oxygen saturation post-procedure.
Synapse has enriched 10 related papers. Consider them for comparative context: