Key result
Simulated atrial arrhythmias increased closing, leakage, and total regurgitant volumes and fractions of a monoleaflet prosthetic heart valve compared to normal sinus rhythm in vitro.
Why the study?
Do simulated atrial arrhythmias increase regurgitation in a monoleaflet prosthetic heart valve compared to normal sinus rhythm?
Do simulated atrial arrhythmias increase regurgitation in a monoleaflet prosthetic heart valve compared to normal sinus rhythm?
In vitro simulation demonstrates that atrial arrhythmias increase regurgitant volumes and fractions in monoleaflet prosthetic heart valves compared to normal sinus rhythm.
Answers are grounded in this paper and cite their sources.
May increase regurgitation risk in monoleaflet valves during AF; hypothesis-generating and leaves open need for clinical validation.
Many patients who receive a prosthetic heart valve also have or acquire cardiac arrhythmias. However, most in vitro studies of prosthetic valves examine them under normal rhythms. In this study, a monoleaflet prosthetic heart valve was tested in vitro under conditions that simulated normal sinus rhythm, first degree atrioventricular heart block, and atrial fibrillation (fixed and variable ventricular rates). Atrial contraction was simulated by an active atrial chamber. The timing between the atrium and ventricle was adjusted to simulate various types of arrhythmias. The closing, leakage, and total regurgitant volumes and fractions increased for each type of atrial arrhythmia when compared to normal sinus rhythm. The peak regurgitant flow increased for first degree atrioventricular heart block and atrial fibrillation with a fixed ventricular rate compared to normal sinus rhythm.
No takes yet. Share an insight, caveat, or question.
Conrad M. Zapanta (2005) studied Prosthetic heart valve with atrial arrhythmias. Simulated atrial arrhythmias (first degree atrioventricular heart block and atrial fibrillation) vs. Normal sinus rhythm was evaluated on Closing, leakage, and total regurgitant volumes and fractions; peak regurgitant flow. Simulated atrial arrhythmias increased closing, leakage, and total regurgitant volumes and fractions of a monoleaflet prosthetic heart valve compared to normal sinus rhythm in vitro.
Synapse has enriched 10 related papers. Consider them for comparative context: