Key result
A pulsatile impeller assist heart and total heart demonstrated good blood compatibility with no severe blood damage or organ dysfunction in calves (up to 11 days) and pigs (up to 6 hours).
Why the study?
Does a pulsatile implantable impeller assist heart and total heart demonstrate acceptable blood compatibility in animal models?
Does a pulsatile implantable impeller assist heart and total heart demonstrate acceptable blood compatibility in animal models?
Answers are grounded in this paper and cite their sources.
The pulsatile impeller assist heart and total heart demonstrated good blood compatibility and no severe organ dysfunction in short-term animal models.
A pulsatile impeller assist heart and a total heart were tested as a chronic left ventricular assist device in 5 calves and an acute biventricular assist device in 4 pigs respectively, to evaluate their blood compatibility. During the left ventricular assist experiments, the indicators for hemolysis, thrombogenesis, renal dysfunction, and hepatic dysfunction were measured preoperatively, at the beginning of the pumping, 6 h postoperatively, and every following day. The results demonstrated that the impeller assist heart causes no severe blood damage nor organ dysfunction in the experiments lasting up to 11 days. In biventricular assist experiments, the number of red blood cells, white blood cells, platelets, and the hematocrit, hemoglobin, free hemoglobin, and lactate dehydrogenase levels were tested preoperatively at the beginning of the pumping and every 2 h postoperatively. The data remained in acceptable ranges during experiments lasting 6 h. It is confirmed that the authors' impeller assist heart and total heart have the advantages of simplicity, implantability, and pulsatility with good blood compatibility.
No takes yet. Share an insight, caveat, or question.
Qian et al. (1995) studied this question. Pulsatile impeller assist heart and total heart was evaluated on Blood compatibility (indicators for hemolysis, thrombogenesis, renal dysfunction, hepatic dysfunction, and blood cell counts). A pulsatile impeller assist heart and total heart demonstrated good blood compatibility with no severe blood damage or organ dysfunction in calves (up to 11 days) and pigs (up to 6 hours).
Synapse has enriched 10 related papers. Consider them for comparative context: