Key result
Assistance with a miniaturized rotary blood pump unloaded the heart, demonstrated by a drop in dP/dt max from 1645 to 1113 mmHg/s.
Why the study?
Does a miniaturized implantable rotary blood pump in an atrial-aortic position restore hemodynamics in a sheep model of heart failure?
Does a miniaturized implantable rotary blood pump in an atrial-aortic position restore hemodynamics in a sheep model of heart failure?
A miniaturized rotary blood pump using atrial cannulation effectively unloads the left ventricle and restores hemodynamics in a sheep model of severe heart failure.
Answers are grounded in this paper and cite their sources.
Supports development of miniaturized atrial-aortic pumps for LV unloading; leaves open translation to human HF trials.
The purpose of this study was to test the pump performance of a miniaturized, newly developed, implantable rotary blood pump (the diagonal pump). In a series of six sheep, the pump was interposed in an atrial-aortic connection. Heart failure was induced by serial injection of glass beads into the left coronary artery. The assistance of the non-failing heart (with pump output up to 4.5 liter/min) did not change the total cardiac output or the blood pressure of the animal. The heart was significantly unloaded as demonstrated by a drop in first derivative of the left ventricular pressure (dP/dt) max (from 1645 to 1113 mmHg/s; P = 0.0003). Because of the specific interaction between heart function and rotary blood pump performance, the pump provided considerably more flow in the failing than in the non-failing heart. Cardiac output and perfusion pressures were restored to pre-assist values in the failing heart model. The heart-rotary blood pump interaction is an ideal substrate for long-term assistance. With this miniaturized rotary blood pump, hemodynamics in a severely failing heart can be restored. Atrial cannulation, which leaves the left ventricle untouched, provides similar results as apical cannulation in the failing heart.
No takes yet. Share an insight, caveat, or question.
Bart Meyns (1998) studied Heart failure (n=6). Miniaturized implantable rotary blood pump (diagonal pump) vs. Baseline (pre-assist values) was evaluated on First derivative of the left ventricular pressure (dP/dt) max (p=0.0003). Assistance with a miniaturized rotary blood pump unloaded the heart, demonstrated by a drop in dP/dt max from 1645 to 1113 mmHg/s.