Key result
Argatroban-loaded polymeric-metallic stents significantly reduced platelet deposition at 2 hours compared to Argatroban-leached stents (1.74 vs 33.5 x 10^8 platelets/cm2; P=0.005) in a swine model.
Why the study?
Does local delivery of Argatroban from a polymeric-metallic stent reduce platelet deposition in a swine coronary model?
RCT (n=8)
Randomly deployed
Does local delivery of Argatroban from a polymeric-metallic stent reduce platelet deposition in a swine coronary model?
Absolute Event Rate: 1.74% vs 33.5%
p-value: p=0.005
Local delivery of Argatroban from a polymer-metallic stent significantly reduces platelet deposition in a swine coronary model.
Supports preclinical argatroban-eluting stent development; leaves open human translation and clinical testing.
Thrombus formation after intracoronary stent implantation provides a stimulus for neointimal hyperplasia and if excessive can result in stent thrombosis. We tested the hypothesis that local delivery of an antithrombin drug from a polymeric-metallic stent inhibits platelet thrombus formation. An uncoated metal slotted tube, a jellyroll slotted metal stent with an Argatroban-loaded polymeric sleeve, and a jellyroll slotted metal stent with a drug-leached polymeric sleeve were randomly deployed into the coronary arteries of eight juvenile farm swine. Platelet deposition in the stented segments was determined at 2 hr using autologous 111Indium oxime-labeled platelets. Platelet deposition was significantly less in the Argatroban-loaded stents compared to the Argatroban-leached stents (1.40 x 10(8) platelets/cm2 vs. 26.8 x 10(8) platelets/cm2; P=0.005). When corrected for differences in the metal surface area exposed to blood, platelet deposition was significantly lower in the Argatroban-loaded stent (1.74+/-1.95 x 10(8)/cm2) compared to the Argatroban-leached stent (33.5+/-39.1 x 10(8)/cm2; P=0.005) and the uncoated metal stent (36.2+/-73.3 x 10(8)/cm2; P=0.006). In this coronary stent thrombosis model Argatroban has local antithrombotic properties when delivered with a polymer-metallic stent. Improved polymeric designs may reduce risk of thrombus deposition at the site of stent implantation.
No takes yet. Share an insight, caveat, or question.
Kruse et al. (1999) conducted an RCT in Thrombus formation after intracoronary stent implantation (n=8). Argatroban-loaded polymeric sleeve on a jellyroll slotted metal stent vs. Argatroban-leached polymeric sleeve on a jellyroll slotted metal stent and uncoated metal slotted tube was evaluated on Platelet deposition in the stented segments corrected for metal surface area exposed to blood (x 10^8 platelets/cm2) (p=0.005). Argatroban-loaded polymeric-metallic stents significantly reduced platelet deposition at 2 hours compared to Argatroban-leached stents (1.74 vs 33.5 x 10^8 platelets/cm2; P=0.005) in a swine model.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: