Key result
Plasmapheresis filters eliminated more recombinant hirudin than hemofilters (60%-70% vs 15%-42%) in less time (80 vs 180 s) during simulated cardiopulmonary bypass.
Why the study?
Do plasmapheresis filter systems improve the elimination of recombinant hirudin compared to hemofilter systems in a simulated cardiopulmonary bypass model?
Population
Closed in vitro simulated cardiopulmonary bypass system testing 30 filters
Comparison
Plasmapheresis filter systems vs Hemofilter systems
Design
Preclinical
Authors
Loading...
Plasmapheresis may speed hirudin removal in CPB; animal model data leave open clinical translation and safety.
Do plasmapheresis filter systems improve the elimination of recombinant hirudin compared to hemofilter systems in a simulated cardiopulmonary bypass model?
In a simulated cardiopulmonary bypass model, plasmapheresis filters provided more rapid and complete elimination of recombinant hirudin compared to standard hemofilters.
Koster et al. (2000) studied Simulated cardiopulmonary bypass (n=30). Plasmapheresis filters vs. Hemofilters was evaluated on Elimination of r-hirudin. Plasmapheresis filters eliminated more recombinant hirudin than hemofilters (60%-70% vs 15%-42%) in less time (80 vs 180 s) during simulated cardiopulmonary bypass.