Key result
An elevated right ventricular pacing threshold ≥ 3.0 V/0.4 ms at 1 week (OR 10.7, p=0.036) and lead bending on chest radiography (OR 12.8, p=0.018) independently predicted chronic His-bundle pacing threshold deterioration.
Why the study?
Concerns about long-term threshold deterioration with His-bundle pacing persist, and predictors of chronic threshold elevation are not yet well established.
What are the early predictors of chronic His-bundle pacing threshold deterioration in patients with successful His-bundle pacing?
Cohort (n=47)
No
What are the early predictors of chronic His-bundle pacing threshold deterioration in patients with successful His-bundle pacing?
Odds Ratio: 10.7
p-value: p=0.036
An elevated RV pacing threshold at 1 week and lead flexion at implantation strongly predict long-term deterioration of His-bundle pacing thresholds, highlighting the need for close monitoring in these patients.
Background His‐bundle pacing (HBP) facilitates physiological ventricular activation. However, concerns about long‐term threshold deterioration persist. The predictors of chronic threshold elevation are not yet well established. Methods Between February 2018 and December 2021, HBP was attempted in 95 patients undergoing pacemaker implantation. Strict success criteria (threshold ≤ 2.0 V/0.4 ms) were applied, and 47 patients with successful HBP were monitored for over 3 years. We assessed pacing thresholds for both the His‐bundle and right ventricle (RV) at implantation, 1 week, 1 month, and annually thereafter. Lead shape was evaluated using chest radiography. Patients were categorized into two groups based on their His‐bundle pacing threshold: stable (≤ 3.5 V/0.4 ms) and deteriorated (> 3.5 V/0.4 ms). Results Deterioration in His‐bundle pacing thresholds was associated with increased RV pacing thresholds. Deterioration in RV pacing occurred earlier, with significant differences observed at 1 week post‐implantation (median RV: 1.87 vs. 3.25 V/0.4 ms, p = 0.032; His‐bundle: 1.0 vs. 1.25 V/0.4 ms, p = 0.212). Multivariate analysis identified an RV threshold ≥ 3.0 V/0.4 ms at 1 week (OR 10.7, p = 0.036) and lead bending on chest radiography (OR 12.8, p = 0.018) as independent predictors of chronic His‐bundle pacing threshold deterioration. Conclusion An elevated RV pacing threshold at 1 week post‐implantation and lead flexion at implantation may serve as early indicators of long‐term deterioration in His‐bundle pacing thresholds. When the RV pacing threshold increase is detected, it is important to closely monitor the patient and frequently adjust the output settings to prevent pacing failure.
No takes yet. Share an insight, caveat, or question.
Sakamoto et al. (2025) conducted a cohort in Pacemaker implantation with successful His-bundle pacing (n=47). Right ventricular pacing threshold ≥ 3.0 V/0.4 ms at 1 week post-implantation vs. Right ventricular pacing threshold < 3.0 V/0.4 ms was evaluated on Chronic His-bundle pacing threshold deterioration (> 3.5 V/0.4 ms) (OR 10.7, p=0.036). An elevated right ventricular pacing threshold ≥ 3.0 V/0.4 ms at 1 week (OR 10.7, p=0.036) and lead bending on chest radiography (OR 12.8, p=0.018) independently predicted chronic His-bundle pacing threshold deterioration.