Describes a method of assessing the counts from 59 Fe in a gate set to detect the 0.32 Mev photopeak of 51 Cr when both isotopes are used together. In the 'single photopeak' procedure comparison is made in the spectra from 59 Fe sources in a water phantom, of the count rate in a gate B set on the 1.1 Mev photopeak and the count rate in a gate C set on the 0.32 photopeak of 51 Cr normalized to the count rate in a gate A set on the 1.29 Mev peak. In the 'double photopeak' procedure comparison is made of the count rate in a gate B, set on the 0.66 Mev photopeak of 137 Cs, with the count rate in gate C, both normalized to the count rate in a gate A set on both the 1.1 and 1.29 Mev photopeaks. It is shown that for certain gate widths there is a linear relationship to within +or-3.3% between the count ratios C/A and B/A which is independent of the size, shape and depth of the 59 Fe source.