Key result
Decreasing transthoracic pulse duration greatly increases the current threshold required to induce VF in dogs.
The current required to induce ventricular fibrillation via transthoracic electroventilation electrodes is significantly higher than that needed for ventilation, especially at shorter pulse durations, suggesting a wide safety margin.
Answers are grounded in this paper and cite their sources.
Supports wide safety margin for short-pulse transthoracic electroventilation in dogs; hypothesis-generating and should not yet change clinical practice.
The threshold for ventricular fibrillation was determined in 12 pentobarbital anesthetized dogs using transthoracic electrodes located at the optimal axillary electroventilation sites. Electroventilation is the name used to designate inspiration produced by stimuli applied to body surface electrodes. The optimal stimulation site for electroventilation was first determined using hand-held electrodes. Then electrodes, 4.1 cm in diameter, were sutured bilaterally to the optimal anterior axillary stimulation site. The threshold current for producing ventricular fibrillation was determined using single pulses ranging from 0.1-10 msec in duration delivered during the vulnerable period of the cardiac cycle. Fibrillation was produced in all dogs with the 10- and 5-msec pulse durations, in 11 dogs with 0.3-msec, in 6 dogs with 0.2-msec, and in 1 dog with 0.1-msec pulse duration. In all dogs, the current required to produce ventricular fibrillation increased greatly as the pulse duration was decreased. The current required for fibrillation was much in excess of that required to produce one tidal volume. With the longer duration pulses, the ratio was about 80. With the 8 microseconds duration pulses used for electroventilation the estimated ratio is about 800.
No takes yet. Share an insight, caveat, or question.
Voorhees et al. (1992) studied Ventricular fibrillation (n=12). Transthoracic electrode stimulation was evaluated on Threshold current for producing ventricular fibrillation. In 12 anesthetized dogs, the current required to produce ventricular fibrillation using transthoracic electrodes increased greatly as the pulse duration was decreased from 10 to 0.1 msec.
Synapse has enriched 10 related papers. Consider them for comparative context: