Key result
Ultrasound echo-tracking with indirect BP underestimates elastic properties by ~28% versus direct arterial measurement.
Why the study?
Does an ultrasound phase-locked echo-tracking system provide reproducible measurements of pulsatile vessel diameter changes and elastic properties in human arteries?
Population
Human subjects
Comparison
Ultrasound phase-locked echo-tracking system… vs Direct intra-arterial blood pressure measurement
Design
Other
Authors
Loading...
Requires caution with indirect BP in echo-tracking stiffness measures; leaves open validation against direct intra-arterial readings.
Does an ultrasound phase-locked echo-tracking system provide reproducible measurements of pulsatile vessel diameter changes and elastic properties in human arteries?
The ultrasound echo-tracking system provides reliable measurements of arterial stiffness, though reliance on indirect blood pressure measurements causes systematic underestimation of elastic modulus and stiffness.
Hansen et al. (1993) studied this question. Ultrasound phase-locked echo-tracking system with indirect blood-pressure measurement vs. Direct intra-arterial blood pressure measurement was evaluated on Reproducibility (intra-observer and inter-observer variability) of pulsatile diameter changes, pressure strain elastic modulus (Ep), and stiffness (beta). The ultrasound echo-tracking system demonstrated intra-observer variabilities of 10-16% for pulsatile diameter changes, but systematically underestimated elastic properties by 25-30% with indirect BP.