Key result
Transapical off-pump mitral valve repair with the NeoChord DS1000 device achieved 100% procedural success and resulted in nonsignificant mitral regurgitation in 81% of patients at 6 months.
Why the study?
To evaluate early results of MV repair using the NeoChord DS1000 device in the first consecutive series of patients operated on in Poland.
Does transapical off-pump mitral valve repair with the NeoChord DS1000 device reduce mitral regurgitation in patients with severe MR due to posterior leaflet prolapse?
Observational (n=21)
No
Does transapical off-pump mitral valve repair with the NeoChord DS1000 device reduce mitral regurgitation in patients with severe MR due to posterior leaflet prolapse?
Transapical off-pump mitral valve repair with the NeoChord DS1000 device is feasible and effectively reduces mitral regurgitation and promotes reverse remodeling at 6 months in selected patients.
Answers are grounded in this paper and cite their sources.
Supports short-term MR reduction with NeoChord in selected prolapse patients; hypothesis-generating, requires randomized trials before practice change.
INTRODUCTION: Minimally invasive techniques of mitral valve (MV) repair have been increasingly used in recent years. Transapical implantation of artificial chordae on a beating heart under 2D/3D transesophageal echocardiographic guidance with the NeoChord DS1000 device is a new surgical treatment of degenerative mitral regurgitation (MR). AIM: To evaluate early results of MV repair with the NeoChord DS1000 device in the first group of consecutive patients operated on in Poland. MATERIAL AND METHODS: Twenty-one patients with severe MR due to posterior leaflet prolapse (81% male; mean age: 60.7 ±12.7 years) underwent MV repair with the NeoChord DS1000 system. There were 12 (57.1%) patients with type A (an isolated central prolapse/flail), 8 (38.1%) patients with type B (multisegment disease/flail) and 1 (4.8%) patient with type C (posterior/paracommisural area) MV prolapse. A flail leaflet was present in 12 (57.1%) patients. The median number of neochords was 3 (2-6). We assessed by echocardiography left-sided heart morphology and evaluated MR degree before and 6 months after chords implantation. RESULTS: Early procedural success was achieved in 100% of patients. At the 6-month follow-up nonsignificant MR (trace and mild) was detected in 17 (81.0%) patients, while moderate MR was detected in 4 (19.0%) patients; mean values of left-sided heart dimensions and volumes, mitral E and E' velocity of lateral MV annulus significantly decreased. CONCLUSIONS: A novel procedure with the NeoChord DS1000 device is feasible in properly selected patients, and results in a significant reduction of MR degree and left ventricle and left atrium reverse remodeling at the 6-month follow-up.
No takes yet. Share an insight, caveat, or question.
Kurnicka et al. (2019) conducted an observational in Severe mitral regurgitation due to posterior leaflet prolapse (n=21). NeoChord DS1000 device was evaluated on Nonsignificant mitral regurgitation (trace or mild) at 6 months. Transapical off-pump mitral valve repair with the NeoChord DS1000 device achieved 100% procedural success and resulted in nonsignificant mitral regurgitation in 81% of patients at 6 months.