Electromagnetic navigation system reaches right heart targets in ~88% of patients, demonstrating feasibility for lead placement.
Why the study?
Is a novel electro-anatomic navigation system feasible and safe for guiding device lead implantation in patients undergoing EP studies or device implantation?
Is a novel electro-anatomic navigation system feasible and safe for guiding device lead implantation in patients undergoing EP studies or device implantation?
A novel electro-anatomic navigation system is feasible and safe for guiding device lead implantation, offering a potential alternative to traditional fluoroscopy-guided techniques.
Answers are grounded in this paper and cite their sources.
Electromagnetic navigation appears feasible for right heart leads; hypothesis-generating and requires larger trials before clinical use.
INTRODUCTION: Fluoroscopy-guided pacing lead placement has well-recognized limitations and risks. We studied the safety and feasibility of using a novel electromagnetic navigation system specifically designed to guide pacemaker and implantable cardioverter defibrillator lead placement. METHODS: Twenty-four patients (mean age of 54±34 years) underwent the study protocol; 16 before electrophysiology study and eight before device implantation. The navigational deflectable sheath assembly was introduced via the subclavian vein and advanced to seven prespecified targets within the right heart chambers. The time taken to reach each target site was measured. RESULTS: All seven prespecified targets were successfully reached by 21 of 24 patients (88%). The total time required to complete the study protocol ranged from 3.21 to 15.25 minutes (average 8.9 minutes), with an associated mean fluoroscopy time of 50±36 seconds. In three of the 24 patients, this navigation system was successfully used to guide right ventricular pacing lead placement. The average total procedure time for these devices was 97.8 minutes (excluding the study protocol), with an average associated fluoroscopy time of 6.93 minutes. These procedures were well tolerated and no periprocedural complications were noted. CONCLUSIONS: This study suggests that this novel electro-anatomic navigation system is a viable and safe alternative to traditional fluoroscopy-guided lead implantation. Further studies are required to determine the absolute reduction in radiation exposure and increased efficiency relative to current standard fluoroscopic techniques.
No takes yet. Share an insight, caveat, or question.
Manlucu et al. (2012) studied Patients undergoing electrophysiology study or device implantation (n=24). Electromagnetic navigation system was evaluated on Successful reaching of seven prespecified targets within the right heart chambers. An electromagnetic navigation system successfully reached all seven prespecified right heart targets in 88% (21 of 24) of patients, demonstrating feasibility for guiding lead placement.