Key result
Complete percutaneous retrieval successfully extracts an embolized M-TEER device from the right subclavian artery.
Why the study?
Complete embolization of an M-TEER device is exceedingly rare, and all previously reported retrievals required final vascular surgery cut down.
Case Report (n=1)
Demonstrates the feasibility of a completely percutaneous approach for retrieving an embolized MitraClip from the subclavian artery, avoiding the need for surgical cutdown.
Answers are grounded in this paper and cite their sources.
May enable nonsurgical management of subclavian M-TEER embolization; extends single-case experience and leaves open broader applicability.
Single leaflet device attachment (SLDA) after mitral transcatheter valve edge-to-edge repair (M-TEER) has been reported in up to 5% of patients. However, complete embolization of the M-TEER device is exceedingly rare. Device retrievals in all previously reported cases were completed by vascular surgery cut down at the end. In this case, we are presenting a complete percutaneous retrieval of M-TEER device from the right subclavian artery, which indicates the importance of procedural planning and appropriate equipment selection.
No takes yet. Share an insight, caveat, or question.
Aalaei‐Andabili et al. (2026) conducted a case report in Embolized MitraClip (M-TEER device) in the right subclavian artery (n=1). Percutaneous retrieval of M-TEER device was evaluated on Successful percutaneous retrieval. Complete percutaneous retrieval of an embolized M-TEER device from the right subclavian artery was successfully performed.