Key result
Toronto Stentless Porcine Valve shows a smaller mean gradient increase during exercise vs Sorin mechanical valves.
Why the study?
Does the Toronto Stentless Porcine Valve improve stress hemodynamics compared to mechanical valves in patients with aortic valve prostheses?
Observational (n=34)
Does the Toronto Stentless Porcine Valve improve stress hemodynamics compared to mechanical valves in patients with aortic valve prostheses?
The Toronto Stentless Porcine Valve demonstrates a smaller increase in transvalvular gradients during dobutamine stress compared to the Sorin Bicarbon mechanical valve, indicating superior hemodynamic performance under stress.
Answers are grounded in this paper and cite their sources.
May support stress hemodynamic advantage for stentless valves; hypothesis-generating and should not yet change prosthesis selection.
BACKGROUND: While aortic valve prostheses are known to perform well at rest, few studies have examined them under stress. We compared stress hemodynamics of mechanical valves and nonstented porcine valves in the aortic position to that of normal native aortic valves. METHODS: Dobutamine echocardiography was used to assess mean and peak gradients and effective orifice area index (EOAI) at rest and exercise in patients with the Toronto Stentless Porcine Valve (SPV) (n = 13, mean implant size 25.7 mm), Sorin Bicarbon mechanical valve (SOR) (n = 11, mean implant size 24.5 mm), and patients with normal native aortic valves (NOR) (n = 10). Dobutamine infusion was started at 5 micron/kg per minute, and increased by increments of 5 micron/kg per minute until the target heart rate was achieved or until a maximal dose of 40 micron/kg per minute. RESULTS: At rest and exercise, respectively, cardiac output (L/min) was 5.2 and 10.4 for Toronto SPV; 7.4 and 13.5 for SOR; and 4.6 and 11.2 for NOR. Measured EOAI (cm2) was 1.1+/-0.2 and 1.15+/-0.2 for TORONTO SPV; 1.60+/-0.3 and 1.58+/-0.3 for SOR; and 1.45+/-0.2 and 1.46+/-0.2 for NOR. Mean gradients (mmHg) were 5.48+/-1.1 and 5.83+/-0.9 for TORONTO SPV; 5.26+/-0.8 and 11.3+/-1.8 for SOR; and 1.54+/-0.4 and 2.18+/-0.7 for NOR. Peak gradients (mmHg) were 11.9+/-2.0 and 21.0+/-3.7 for TORONTO SPV; 10.79+/-1.7 and 25.9+/-3.4 for SOR; and 2.38+/-0.9 and 6.1+/-2.3 for NOR. CONCLUSIONS: Although the mechanical group (SOR) had larger measured EOAI, the greater increase in gradients with exercise in this group suggests that the TORONTO SPV is less obstructive to flow.
No takes yet. Share an insight, caveat, or question.
Shaheen et al. (1998) conducted an observational in Aortic valve prostheses (n=34). Toronto Stentless Porcine Valve vs. Sorin Bicarbon mechanical valve and normal native aortic valves was evaluated on Mean and peak gradients and effective orifice area index (EOAI) at rest and exercise. During dobutamine exercise, the Toronto Stentless Porcine Valve showed a smaller increase in mean gradients (5.48 to 5.83 mmHg) compared to the Sorin mechanical valve (5.26 to 11.3 mmHg).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: