Key result
TactiCath SE ablation yields ~62% freedom from persistent AF recurrence at fifteen months.
Why the study?
Safety and effectiveness of a novel magnetic sensor enabled optical contact force ablation catheter had been established for paroxysmal AF, but had yet to be demonstrated in persistent AF.
Does contact force catheter ablation using TactiCath SE improve freedom from arrhythmia and quality of life in patients with persistent atrial fibrillation refractory to antiarrhythmic drugs?
Does contact force catheter ablation using TactiCath SE improve freedom from arrhythmia and quality of life in patients with persistent atrial fibrillation refractory to antiarrhythmic drugs?
Contact force catheter ablation using the TactiCath SE catheter is safe and effective for treating drug-refractory persistent AF, achieving 61.6% freedom from arrhythmia at 15 months and significantly improving quality of life.
Supports TactiCath SE ablation safety in persistent AF; leaves open randomized confirmation of arrhythmia freedom and QoL gains.
INTRODUCTION: Use of a novel magnetic sensor enabled optical contact force ablation catheter has been established to be safe and effective for treatment of symptomatic drug-refractory paroxysmal atrial fibrillation (AF) but has yet to be demonstrated in the persistent AF (PersAF) population. METHODS: PERSIST-END was a multicenter, prospective, nonrandomized, investigational study designed to demonstrate the safety and effectiveness of TactiCath™ Ablation Catheter, Sensor Enabled™(SE) (TactiCath SE) for use in the treatment of subjects with documented PersAF refractory or intolerant to at least one Class I/III AAD. The ablation strategy included pulmonary vein isolation and additional targets at physician discretion. Follow-up through 15-months, including a 3-month blanking period and 3-month therapy consolidation period, was performed with cardiac event and Holter monitoring. Primary safety, primary effectiveness, clinical success, and quality of life (QOL) endpoints were analyzed. RESULTS: Of 224 subjects enrolled at 21 investigational sites in the United States and Australia, 223 underwent ablation with the investigational catheter. The primary safety event rate was 3.1% (seven events in seven subjects). The Kaplan-Meier estimate of freedom from AF/atrial flutter/atrial tachycardia recurrence at 15-months was 61.6% and clinical success at 15 months was 89.8%. Subject QOL significantly improved following ablation as assessed via AFEQT (31.6 point increase, p < .0001) and EQ-5D-5L (10.7 point increase, p < .0001) and was met with a 53% reduction in all cause cardiovascular healthcare utilization. CONCLUSION: The sensor-enabled force-sensing catheter is safe and effective for the treatment of drug refractory recurrent symptomatic PersAF, reducing arrhythmia recurrence while improving QOL and healthcare utilization.
No takes yet. Share an insight, caveat, or question.
Lo et al. (2022) studied persistent atrial fibrillation (PersAF) (n=224). TactiCath Ablation Catheter, Sensor Enabled was evaluated on freedom from AF/atrial flutter/atrial tachycardia recurrence at 15-months. Ablation with the TactiCath Sensor Enabled catheter for persistent atrial fibrillation yielded a 3.1% primary safety event rate and 61.6% freedom from arrhythmia recurrence at 15 months.