Answers are grounded in this paper and cite their sources.
Introduction Jaundice is the visible manifestation of chemical bilirubinemia and is characterized by yellow discoloration of skin. In a pediatric emergency department, evaluation of jaundice includes a precise medical history, physical examination, and estimation of bilirubin by laboratory tests. Measurement of total serum bilirubin (TSB) by transcutaneous bilirubin (TcB) is interpreted according to the age (in hours) of the newborn infant. Studies have also indicated that the TcB determination is more accurate than visual estimation methods. Aim To compare BiliCare System against standard laboratory testing for bilirubin estimation. Results The coefficient of correlation was r = 0.7245, and the standard deviation between the mean values of TcB and TSB is ±0.55. Conclusion It is concluded from the study that there is good correlation between TcB measurement and laboratory estimation of total bilirubin levels. It is recommended that transcutaneous measurement can be used for initial screening of jaundice as well as monitoring of phototherapy in neonates. How to cite this article Warade J. Comparison of BiliCare Transcutaneous Bilirubinometer with Standard Laboratory Assay. Indian J Med Biochem 2017;21(1):1-4.
No takes yet. Share an insight, caveat, or question.
Jayesh Warade (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: