Moor MBF3D laser Doppler correlates well with the Perimed PF3 in an in vitro model.
Why the study?
A new infrared laser Doppler blood flow instrument (Moor MBF3D) required evaluation against an established device over a range of flow velocities and concentrations.
Answers are grounded in this paper and cite their sources.
The Moor MBF3D infrared laser Doppler blood flowmeter provides comparable measurements to the Perimed PF3 in an in vitro model, with the added advantage of dual-channel operation for simultaneous multi-site measurements.
A new infrared laser Doppler blood flow instrument (Moor MBF3D) was evaluated using an in vitro model allowing measurements over a range of flow velocities and concentrations. The responses correlated well (r = 0.96, p less than 0.01) with those obtained simultaneously using a Perimed PF3 laser Doppler instrument. The different processing bandwidths of the instruments were investigated and the wideband mode of operation is recommended for flow measurements where there may be fast moving red blood cells (rbcs). The infrared instrument is capable of dual-channel operation, and the two channels are shown to respond almost identically for similar changes in blood flow through the in vitro model (r = 0.999, p less than 0.01). The main advantage of the dual-channel instrument is that continuous measurements may be made simultaneously at two different skin sites allowing dynamic flow responses to be compared.
No takes yet. Share an insight, caveat, or question.
Barnett et al. (1990) studied this question. Moor MBF3D infrared laser Doppler blood flow instrument vs. Perimed PF3 laser Doppler instrument was evaluated on Flow velocity and concentration measurement responses (r = 0.96, p=<0.01). The Moor MBF3D infrared laser Doppler blood flow instrument correlated well with the Perimed PF3 instrument in an in vitro model.
Synapse has enriched 10 related papers. Consider them for comparative context: