Key result
Computerized ultrasonic arteriography detects carotid narrowing with 100% sensitivity compared to vessel X-rays.
Why the study?
Does computerized ultrasonic arteriography accurately detect carotid occlusive disease compared to vessel X-rays?
Observational (n=94)
Does computerized ultrasonic arteriography accurately detect carotid occlusive disease compared to vessel X-rays?
Computerized ultrasonic arteriography is a highly sensitive noninvasive technique for detecting significant carotid artery stenosis.
Answers are grounded in this paper and cite their sources.
May support noninvasive carotid screening; leaves open prospective validation before replacing angiography.
The development of a new noninvasive computerized technique for producing simultaneous lateral, anterior-posterior, and transverse views of the carotid bifurcation for the diagnosis of occlusive disease is described. These multiplanar views have been used to detect 40 percent diameter reductions with a sensitivity of 100 percent and a specificity of 85 percent in 94 comparisons with vessel X-rays.
No takes yet. Share an insight, caveat, or question.
Miles et al. (1982) conducted an observational in Carotid occlusive disease (n=94). Computerized ultrasonic arteriography vs. Vessel X-rays was evaluated on Detection of 40 percent diameter reductions (Sensitivity 100%, Specificity 85%). Computerized ultrasonic arteriography detected 40% diameter reductions in the carotid bifurcation with 100% sensitivity and 85% specificity compared to vessel X-rays.
Synapse has enriched 10 related papers. Consider them for comparative context: