Key result
Balloon coronary testing was classified as inaccurate in 36.1% of cases during percutaneous pulmonary valve implantation, indicating it may not accurately predict final valve position.
Why the study?
Coronary compression has occurred after pulmonary valve implantation despite a negative balloon coronary test, calling the test's reliability and accuracy into question.
Does balloon coronary compression testing accurately predict final valve position during percutaneous pulmonary valve implantation?
Observational (n=36)
Does balloon coronary compression testing accurately predict final valve position during percutaneous pulmonary valve implantation?
Balloon coronary testing during percutaneous pulmonary valve implantation may be inaccurate in predicting final valve position, potentially leading to a high false positive rate for coronary compression risk.
Answers are grounded in this paper and cite their sources.
May overestimate coronary compression risk during PPVI; leaves open validation of alternative predictors in prospective studies.
OBJECTIVE: To evaluate the reliability of balloon coronary compression testing during percutaneous pulmonary valve implantation. BACKGROUND: Despite the widespread use of the 'balloon coronary test' as the preferable method to rule out the risk of coronary compression, this adverse event has been described after pulmonary valve implantation where coronary balloon test suggested no risk or low risk, calling into question the accuracy of the test. METHODS: We performed a retrospective chart review of 84 patients who underwent pulmonary valve implantation between January 2018 and December 2019 and selected 36 patients whose archived imaging was suitable to perform quantitative analysis of the 'balloon coronary test'. We focused on the spatial disparity between the right ventricular outflow tract position defined by the inflated testing balloon and the eventual implanted valve position, to classify the test as inaccurate or accurate. RESULTS: In total, 36.1% of cases were classified as having an inaccurate coronary balloon test. Among the baseline characteristics, right ventricular outflow tract substrate was identified as a significant predictor of test accuracy. Related to this characteristic, the type of testing balloon used and the size of the eventually implanted valve were found to be associated with test accuracy. CONCLUSIONS: Based on our findings, balloon coronary testing is not an accurate method of predicting final valve position with respect to fixed structures in the thorax. This may translate to a high false positive rate for the likelihood of coronary compression in pulmonary valve implantation.
No takes yet. Share an insight, caveat, or question.
Morgan et al. (2021) conducted an observational in Percutaneous pulmonary valve implantation (n=36). Balloon coronary compression testing was evaluated on Inaccurate coronary balloon test (spatial disparity between the right ventricular outflow tract position defined by the inflated testing balloon and the eventual implanted valve position). Balloon coronary testing was classified as inaccurate in 36.1% of cases during percutaneous pulmonary valve implantation, indicating it may not accurately predict final valve position.