Key result
Image-guided focused ultrasound achieves non-invasive cardiac pacing in pigs using contrast agents to overcome attenuation.
Why the study?
No non-invasive cardiac pacing device acceptable for prolonged use in conscious patients currently exists.
Does High Intensity Focused Ultrasound (HIFU) enable effective and safe non-invasive cardiac pacing in a preclinical pig model?
Does High Intensity Focused Ultrasound (HIFU) enable effective and safe non-invasive cardiac pacing in a preclinical pig model?
Answers are grounded in this paper and cite their sources.
High Intensity Focused Ultrasound (HIFU) demonstrates feasibility and acute safety for non-invasive cardiac pacing in a preclinical pig model.
Currently, no non-invasive cardiac pacing device acceptable for prolonged use in conscious patients exists. High Intensity Focused Ultrasound (HIFU) can be used to perform remote pacing using reversibility of electromechanical coupling of cardiomyocytes. Here we described an extracorporeal cardiac stimulation device and study its efficacy and safety. We conducted experiments ex vivo and in vivo in a large animal model (pig) to evaluate clinical potential of such a technique. The stimulation threshold was determined in 10 different ex vivo hearts and different clinically relevant electrical effects such as consecutive stimulations of different heart chambers with a single ultrasonic probe, continuous pacing or the inducibility of ventricular tachycardia were shown. Using ultrasonic contrast agent, consistent cardiac stimulation was achievable in vivo for up to 1 hour sessions in 4 different animals. No damage was observed in inversion-recovery MR sequences performed in vivo in the 4 animals. Histological analysis revealed no differences between stimulated and control regions, for all ex vivo and in vivo cases.
No takes yet. Share an insight, caveat, or question.
Marquet et al. (2016) studied Normal cardiac function (animal model) (n=14). Image-guided focused ultrasound (HIFU) vs. Baseline sinus rhythm / unsonicated control regions was evaluated on Acoustic threshold for reliable cardiac stimulation (≥90% success rate). Image-guided focused ultrasound successfully achieved non-invasive cardiac pacing in a pig model, with consistent in vivo stimulation requiring the use of ultrasound contrast agents to overcome acoustic attenuation.