Key result
Sofia FIA detects influenza A with 82% sensitivity but false positives limit influenza B utility.
Why the study?
The diagnostic performance of the Sofia influenza A+B fluorescent immunoassay in adult outpatients with influenza-like illness was evaluated to determine its accuracy compared to existing tests.
Does the Sofia influenza A+B fluorescent immunoassay accurately detect influenza A and B viruses in adult outpatients with influenza-like illness?
Observational (n=109)
No
Does the Sofia influenza A+B fluorescent immunoassay accurately detect influenza A and B viruses in adult outpatients with influenza-like illness?
The Sofia FIA effectively detects influenza A virus, especially with nasopharyngeal samples, but has a high false positive rate for influenza B.
May aid influenza A detection but not B in adult outpatients; leaves open its clinical role pending larger validation studies.
To evaluate the diagnostic performance of the Sofia influenza A+B fluorescent immunoassay (Sofia FIA), we performed a prospective study at the Chang Gung Memorial Hospital in Taiwan from January 2012 to December 2013. Patients who presented at out-patient clinics or the emergency department with influenza-like illness were included. Upper respiratory tract specimens were collected from oropharynx or nasopharynx. Performance of the Sofia FIA was compared to that of the Formosa One Sure Flu A/B Rapid Test. A Real-time reverse transcriptase-polymerase chain reaction assay (RT-PCR) and/or virus culture were used as reference standards. Of the 109 enrolled patients, the sensitivity, specificity, positive, and negative predictive values of the Sofia FIA to detect influenza A virus were 82%, 89%, 77%, and 89%, respectively. These parameters were 100% when the samples were from nasopharynx. The positive predictive value for influenza B virus detection was 29%. The sensitivity of the Sofia FIA for detection of influenza A virus was 93% between days 2 and 4 after onset of symptoms. For specimens with low viral loads (RT-PCR cycle threshold between 30 and 34.9), the sensitivity of The Sofia FIA was 83% (10/12). The Sofia FIA performed effectively in detecting influenza A virus infection. With nasopharyngeal samples, the performance was comparable to RT-PCR. Although influenza viral load typically decreases with time, the Sofia FIA was sensitive enough to identify influenza infecting patients presenting after several days of illness. However, a high false positive rate limits the assay's usefulness to identify influenza B virus infection.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2018) conducted an observational in Influenza-like illness (n=109). Sofia influenza A+B fluorescent immunoassay (Sofia FIA) vs. Formosa One Sure Flu A/B Rapid Test and RT-PCR/virus culture was evaluated on Sensitivity for detection of influenza A virus. The Sofia FIA demonstrated 82% sensitivity and 89% specificity for detecting influenza A, but a high false positive rate limited its utility for identifying influenza B (PPV 29%).