Key result
Balloon anchor technique safely treats stenotic lesions through prior multilayer stent mesh in 3 patients.
Why the study?
Concerns exist whether the mesh of previously implanted multilayer stents can be crossed with existing interventional devices to treat de novo stenotic lesions in covered aortic side branches.
Can the mesh of previously implanted multilayer stents be safely crossed with interventional devices to treat stenotic lesions in covered aortic side branches?
Case Report (n=3)
Can the mesh of previously implanted multilayer stents be safely crossed with interventional devices to treat stenotic lesions in covered aortic side branches?
This case series demonstrates the feasibility and safety of crossing previously implanted multilayer stents using a balloon anchor technique to treat side-branch stenosis endovascularly, avoiding open surgery.
Answers are grounded in this paper and cite their sources.
Endovascular crossing of multilayer stent mesh for side branch stenoses appears feasible in select cases; leaves open questions of safety and durability.
Concerns exist whether the mesh of previously implanted multilayer stents can be crossed with existing interventional devices to treat endovascularly de novo stenotic lesions in covered aortic side branches. We describe the endovascular procedures used to treat three symptomatic patients (two male, one female; mean age: 69 years) who had undergone treatment for thoracoabdominal aortic aneurysms with multilayer stents a mean of 40 months earlier. The procedure involved six general steps: 1) using a brachial approach to navigate a 6 Fr Judkins Right guiding catheter close to the target branch ostium; 2) crossing the stent’s mesh covering the targeted side-branch ostium with a high-torque soft-tip guidewire; 3) advancing a low-profile, semicompliant 2-mm coronary balloon across the mesh and dilating it; 4) exchanging the 2-mm for a 3- or 4-mm noncompliant coronary balloon, and applying the balloon anchor technique to advance the guiding catheter across the mesh to engage it in the branch artery; 5) treating the stenotic lesion with balloon predilatation; and 6) implanting a low-profile, high-radial force stent through the mesh with the side branch stent protruding into the aorta. All target lesions were successfully treated without complications and patent at six-month follow-up. This case series demonstrates that the mesh of previously implanted multilayer stents can be crossed safely with readily available interventional devices using the balloon anchor technique. Endovascular treatment of multilayer stent-covered visceral and renal arteries is possible without the need for open surgery. These findings should be validated in larger studies with longer follow-up.
No takes yet. Share an insight, caveat, or question.
Petrov et al. (2025) conducted a case report in Ostial stenosis in aortic side branches covered by multilayer stents (n=3). Endovascular treatment crossing multilayer stent mesh using balloon anchor technique was evaluated on Successful treatment without complications and patency. Endovascular treatment crossing the mesh of previously implanted multilayer stents using the balloon anchor technique successfully treated stenotic lesions in all 3 patients without complications.
Synapse has enriched 10 related papers. Consider them for comparative context: