Key result
Simulated compliant pulmonary visceral pleura aortic valves yield higher stroke volumes and lower leaflet stress.
Why the study?
The impact of aortic valve leaflet bending rigidity on valve performance and stress distribution has not been well studied, particularly comparing pulmonary visceral pleura and pericardial tissue.
Does pulmonary visceral pleura tissue improve aortic valve performance and reduce leaflet stress compared to pericardial tissue in a computational model?
Does pulmonary visceral pleura tissue improve aortic valve performance and reduce leaflet stress compared to pericardial tissue in a computational model?
Computational modeling suggests that pulmonary visceral pleura, due to its higher compliance, may offer superior hemodynamic performance and lower systolic leaflet stress compared to standard pericardial tissue for bioprosthetic heart valves.
Flexible biomaterials merit preclinical testing in valve prototypes; leaves open in vivo durability and clinical translation.
Background: Porcine heart and bovine pericardium valves, which are collagen-based with relatively little elastin, have been broadly utilized to construct bioprosthetic heart valves (BHVs). With a larger proportion of elastin, the pulmonary visceral pleura (PVP) has greater elasticity and could potentially serve as an advantageous biomaterial for the construction/repair of BHVs. The question of how the aortic valve’s performance is affected by its bending rigidity has not been well studied. Methods: Based on the stress–strain relationships of the pericardium and PVP determined by planar uni-axial tests, a three-dimensional (3D) computational fluid–structure interaction (FSI) framework is employed to numerically investigate the aortic valve’s performance by considering three different cases with Young’s modulus as follows: E=375 , 750 , and 1500 kPa, respectively. Results: The stroke volumes are 112, 99.6, and 91.4 ml as Young’s modulus increases from 375 to 750 and 1500 kPa, respectively. Peak geometric opening area (GOA) values are 2.3, 2.2, and 2.0 cm2 for E=375 , 750, and 1500 kPa, respectively. The maximum value of the aortic leaflet stress is about 271 kPa for E=375 kPa, and it increases to about 383 and 540 kPa for E=750 and 1500 kPa in the belly region at the peak systole, while it reduces from 550 kPa to 450 and 400 kPa for E=375 , 750, and 1500 kPa, respectively, at the instant of peak “water-hammer”. Conclusion: A more compliant PVP aortic leaflet valve with a smaller Young’s modulus, E , has a higher cardiac output, larger GOA, and lower hemodynamic resistance. Most importantly, the aortic leaflet stresses are substantially lower in the belly region within the higher compliance PVP aortic valve tissue during the systole phase, even though some stress increase is also found during the fast-closing phase due to the “water-hammer” effect similar to that in the pericardial tissue. Future clinical studies will be conducted to test the hypothesis that the PVP-based valve leaflets with higher compliance will have lower fatigue or calcification rates due to the overall lower stress.
No takes yet. Share an insight, caveat, or question.
Lu et al. (2022) studied Bioprosthetic heart valves. Pulmonary visceral pleura (PVP) tissue (E=375 kPa) vs. Pericardial tissue (E=750 and 1500 kPa) was evaluated on Stroke volume, geometric opening area (GOA), and aortic leaflet stress. A simulated pulmonary visceral pleura aortic valve with lower Young's modulus (375 kPa) yielded higher stroke volume (112 ml) and lower peak systolic leaflet stress (271 kPa) than stiffer tissues.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: