Answers are grounded in this paper and cite their sources.
Polycrystalline mercuric iodide nuclear radiation detectors have been produced in a novel technology. Unlike the normal single-crystal technology, there is no intrinsic limit to the surface on which these detectors can be produced. Detectors with areas up to about 1.5 cm/sup 2/, thicknesses from 30 to 600 /spl mu/m, and with single electrodes as well as microstrip and pixel contacts have been fabricated and successfully tested with photons in the range of 40-660 keV, /spl beta/ particle's emitted from a Sr-Y source, and high energy (100 GeV) muons. Results on both charge collection and counting efficiency are reported as well as some very preliminary imaging results. The experimental results on charge collection have been compared with simulation, and a combined /spl mu//spl tau/ product 10/sup -7/ cm/sup 2//V for electrons has been estimated.
No takes yet. Share an insight, caveat, or question.
Schieber et al. (1997) studied this question.