Key result
Porcine aortic roots expand ~40% during pressurization, suggesting unstented xenograft valves require substantial oversizing.
Why the study?
Proper sizing of aortic allografts and stentless porcine xenograft valves for the recipient aortic root is critical but not well characterized in terms of dynamic aortic root expansion.
How does the aortic root expand under varying pressures in a porcine model, and what are the implications for bioprosthetic valve sizing?
How does the aortic root expand under varying pressures in a porcine model, and what are the implications for bioprosthetic valve sizing?
Aortic root expansion is highly pressure-dependent, suggesting unstented xenograft valves should be sized 30% to 40% larger than the collapsed recipient aorta.
Answers are grounded in this paper and cite their sources.
Porcine model enables dynamic aortic root measurements; leaves open translation to human allograft sizing.
The implantation of aortic allografts as well as stentless, freehand porcine xenograft valves requires proper sizing of the graft for the recipient aortic root. To visualize the aortic valve in motion and measure the cyclic expansion of the aortic root, we developed an isolated porcine heart model and a computerized three-dimensional reconstruction technique. Dynamic and static expansions of the aortic root were obtained from beating and arrested porcine hearts, and additional static expansions at varying pressures were measured from reconstructed three-dimensional models of valves obtained with high-resolution magnetic resonance imaging. Measurements of aortic root expansion have shown that it is highly dependent upon the pressures imposed on the heart. Although the aortic root expanded by only 5% between systolic pressures of 60 and 100 mmHg, the total expansion was up to 40% between rest and cyclic pressurizing to 100 mmHg. This data suggest that unstented xenograft valves should be sized 30% to 40% larger than the collapsed size of the recipient aorta. Proper sizing of valves on stents should also be attempted to reduce the large amount of leaflet redundancy that current stenting techniques produce.
No takes yet. Share an insight, caveat, or question.
Veselý et al. (1991) studied Bioprosthetic valve sizing. Cyclic pressurizing vs. Rest was evaluated on Aortic root expansion. Aortic root expansion in porcine hearts reached up to 40% between rest and cyclic pressurizing to 100 mmHg, suggesting unstented xenograft valves should be sized 30% to 40% larger than the collapsed aorta.