Key result
Custom pacemaker alert electrode successfully warns patients via pectoral muscle stimulation at ~1.2 V.
Why the study?
Malfunction of a pacing system can be life-threatening for pacemaker-dependent patients, and a built-in automatic warning system to alert patients of dangerous disorders is desirable.
Does a built-in warning system with a custom-made alert electrode effectively alert pacemaker-dependent patients?
Does a built-in warning system with a custom-made alert electrode effectively alert pacemaker-dependent patients?
A custom-made alert electrode can successfully stimulate the pectoral muscle to alert pacemaker patients at low voltage thresholds.
Answers are grounded in this paper and cite their sources.
Alert electrode feasible in small pilot; leaves open whether automatic warnings improve safety in pacemaker-dependent patients.
Malfunction of a pacing system can be life-threatening for a pacemaker-dependent patient. It would be desirable for implantable pulse generators to have a built-in automatic warning system capable of alerting the patient as soon as a potentially dangerous disorder is detected. In this study, seven patients (mean age, 72.6 +/- 10.7 years) with slow, chronic atrial fibrillation underwent implantation of a dual-chamber pulse generator with a custom-made "alert electrode" connected to the atrial port of the pulse generator to stimulate the underlying pectoral muscle. The muscle was temporarily stimulated while the pacemaker was in VVIR mode. The lowest amplitude sufficient to alert the patient (perception threshold) was 1.6 +/- 0.58 V at 0.45 ms during implantation and 1.2 +/- 0.5 V at 0.45 ms chronically. In a second phase, alerts outside of the office were issued using a special software routine capable of delivering stimuli at programmable date and time.
No takes yet. Share an insight, caveat, or question.
Fåhraeus et al. (1998) studied slow, chronic atrial fibrillation (n=7). Dual-chamber pulse generator with a custom-made alert electrode was evaluated on Lowest amplitude sufficient to alert the patient (perception threshold). A custom-made alert electrode connected to a pacemaker successfully alerted patients via pectoral muscle stimulation, with a mean chronic perception threshold of 1.2 +/- 0.5 V at 0.45 ms.