Key result
Impedance cardiography correlates with dye-dilution for stroke index but yields lower values and unreliable changes.
Why the study?
The reliability of impedance cardiography for determining absolute values and changes in stroke index compared to the dye-dilution technique was uncertain.
Does impedance cardiography reliably determine stroke index compared to the dye-dilution technique in human subjects?
Population
17 subjects with 122 stroke index determinations
Comparison
Impedance cardiography vs dye-dilution technique
Design
Observational study
Authors
Loading...
Impedance cardiography is unreliable for stroke index; challenges clinical adoption and leaves open validation of alternatives.
Observational (n=17)
Does impedance cardiography reliably determine stroke index compared to the dye-dilution technique in human subjects?
Effect estimate: r=0.49
p-value: p=<0.001
Impedance cardiography does not reliably determine absolute values of stroke index or evaluate changes compared to the dye-dilution technique.
Keim et al. (1976) reported an observational. Impedance cardiography vs. Dye-dilution technique was evaluated on Stroke index determination (r=0.49, p=<0.001). Impedance cardiography for determining stroke index correlated with the dye-dilution technique (r=0.49, P<0.001) but yielded significantly lower absolute values and unreliable changes within subjects.