Key result
Virtual high radial force TAVI devices reduce paravalvular area vs high conformability devices in ~75% of BAVs.
Why the study?
Bicuspid aortic valve patients have complex valve anatomy often contraindicating TAVI, and the comparative performance of high conformability versus high radial force devices in this group is unclear.
Do high radial force TAVI devices improve paravalvular orifice area and stent-root contact area compared to high conformability devices in patients with bicuspid aortic valves?
Do high radial force TAVI devices improve paravalvular orifice area and stent-root contact area compared to high conformability devices in patients with bicuspid aortic valves?
Computational simulations suggest that high radial force TAVI devices may provide better annular reshaping and lower paravalvular leak than high conformability devices in highly calcified bicuspid aortic valves.
Answers are grounded in this paper and cite their sources.
May favor high radial force TAVI in bicuspid valves; hypothesis-generating computational data leave clinical translation open.
OBJECTIVE: Transcatheter Aortic Valve Implantation (TAVI) is a consolidated procedure showing a low operative risk and excellent long-term outcomes in patients with aortic stenosis. Patients presenting a bicuspid aortic valve (BAV) often require valve replacement due to the highly calcific nature of the aortic leaflets. However, BAV patients have usually been contraindicated for TAVI due to their complex valve anatomy. The aim of this work was to compare the performance of devices featuring high conformability (HC) against those with high radial force (HRF). METHODS: Four BAV patients undergoing TAVI were retrospectively selected. The aortic roots including the native leaflets and calcifications were reconstructed from pre-operative Computed Tomography scans. In each patient, both HC and HRF devices were virtually implanted using Finite Element Analysis simulations. After implantation, paravalvular orifice area, von Mises stress distribution, root contact area, and device eccentricity were calculated. RESULTS: Simulations showed good agreement with intraoperative imaging. In 3 out of 4 patients, the HRF device resulted in a lower paravalvular area than the HC. Stress distribution was also more homogeneously distributed in the HRF group as compared with the HC group. Despite their lower adaptability, HRF devices showed consistently higher stent-root contact area. CONCLUSION: HRF devices showed improved results with respect to HC valves after being deployed in BAV anatomies. We hypothesize that the ability to reshape the annulus is the major determinant of success in this subgroup of patients featuring highly calcified leaflets.
No takes yet. Share an insight, caveat, or question.
Finotello et al. (2021) studied Bicuspid aortic valve (BAV) undergoing TAVI (n=4). High radial force (HRF) devices vs. High conformability (HC) devices was evaluated on Paravalvular orifice area, von Mises stress distribution, root contact area, and device eccentricity. Virtual implantation of high radial force devices resulted in a lower paravalvular area than high conformability devices in 3 out of 4 patients with bicuspid aortic valve.