Key result
Stored and real-time electrogram analysis correctly diagnoses inappropriate ICD therapies triggered by rate-dependent RBBB.
Why the study?
Inappropriate ICD therapies occurring during periods of rate-dependent right bundle branch block require detailed analysis for correct diagnosis and management.
Analysis of stored and real-time ICD electrograms is critical for diagnosing and resolving inappropriate ICD therapies caused by rate-related bundle branch block.
Answers are grounded in this paper and cite their sources.
Detailed electrogram review aids troubleshooting of inappropriate ICD therapies in paced patients with conduction block; single-case data leaves optimal strategies open.
A patient with an implantable cardioverter defibrillator (ICD) and a dual chamber pacemaker experienced inappropriate ICD therapies only during periods of rate-dependent right bundle branch block. Analysis of both stored and real-time ICD electrograms was critical to correctly diagnosing the problem and offering a solution.
No takes yet. Share an insight, caveat, or question.
Coyne et al. (1997) studied this question. Analysis of stored and real-time electrograms correctly diagnosed inappropriate implantable cardioverter defibrillator therapies triggered by rate-dependent right bundle branch block.