Inappropriate ICD discharges during exercise can result from intermittent atrial undersensing due to decreased P wave amplitude, correctable by increasing atrial sensitivity.
Answers are grounded in this paper and cite their sources.
May explain exercise-induced inappropriate ICD shocks via atrial undersensing; case report leaves open whether sensitivity adjustments warrant routine consideration.
We present the case of a patient with a dual chamber implantable cardioverter defibrillator (ICD) who experienced inappropriate ICD discharges during exercise. Interrogation of the ICD revealed intermittent atrial undersensing during exercise that was responsible for the erroneous classification by the ICD of sinus tachycardia as ventricular tachycardia. Monitoring of the intracardiac electrograms and Marker Channels during an exercise test confirmed a marked decrease in P wave amplitude during exercise. By increasing the atrial sensitivity setting the problem was resolved.