Why the study?
AV Search+ combined with sensor rate pacing can cause short-coupled atrial paced intervals that may be proarrhythmic, leading to recurrent atrial arrhythmias.
Do device programming changes reduce automatic mode switch burden in a patient with AV hysteresis-induced atrial arrhythmias?
Case Report (n=1)
Do device programming changes reduce automatic mode switch burden in a patient with AV hysteresis-induced atrial arrhythmias?
AV Search+ combined with sensor rate pacing can cause short-coupled atrial paced intervals leading to proarrhythmic events, which can be resolved with appropriate programming changes.
A 73-year-old male with dual-chamber implantable cardioverter defibrillator (Teligen, Boston Scientific, Marlborough, MA, USA) had multiple episodes of automatic mode switch (AMS) during clinical follow-up. Over 50% of these demonstrated a similar pattern of initiation. AV Search+ in combination with sensor rate pacing can cause short-coupled atrial paced intervals that can be proarrhythmic. After programming changes were made AMS burden has significantly decreased.
No takes yet. Share an insight, caveat, or question.
Monkhouse et al. (2018) conducted a case report in Recurrent atrial arrhythmias (n=1). Programming changes was evaluated on Automatic mode switch (AMS) burden.