Key result
Post-PCI optical flow ratio correlates strongly with fractional flow reserve in stented coronary arteries.
Why the study?
The relationship between optical flow ratio (OFR) and fractional flow reserve (FFR) in stent-treated arteries immediately after PCI was not established.
Does Optical Flow Ratio (OFR) accurately correlate with Fractional Flow Reserve (FFR) in stent-treated arteries immediately after PCI?
Observational (n=103)
Single-blind
No
Does Optical Flow Ratio (OFR) accurately correlate with Fractional Flow Reserve (FFR) in stent-treated arteries immediately after PCI?
Answers are grounded in this paper and cite their sources.
Optical flow ratio (OFR) demonstrates high correlation and agreement with invasive FFR for assessing coronary physiology immediately post-PCI, offering a potential wire-free alternative.
BACKGROUND: Optical flow ratio (OFR) is a recently developed method for functional assessment of coronary artery disease based on computational fluid dynamics of vascular anatomical data from intravascular optical coherence tomography (OCT). The purpose of this study was to investigate the relationship between OFR and fractional flow reserve (FFR) in stent-treated arteries immediately after percutaneous coronary intervention (PCI). METHODS AND RESULTS: The OFR and FFR were measured in 103 coronary arteries immediately after successful PCI with a stent. An increase in the OFR and FFR values within the stent was defined as in-stent ∆OFR and ∆FFR, respectively. The values of FFR and OFR were 0.89±0.06 and 0.90±0.06, respectively. OFR was highly correlated with FFR (r=0.84, P<0.001). OFR showed a good agreement with FFR, presenting small values of mean difference and root-mean-squared deviation (FFR-OFR: -0.01±0.04). In-stent ∆OFR showed a moderate correlation (r=0.69, P<0.001) and good agreement (in-stent ∆FFR - in-stent ∆OFR: 0.00±0.02) with in-stent ∆FFR. CONCLUSIONS: OFR showed a high correlation and good agreement with FFR in stent-treated arteries immediately after PCI.
No takes yet. Share an insight, caveat, or question.
Emori et al. (2020) conducted an observational in Coronary artery disease requiring percutaneous coronary intervention (n=103). Optical flow ratio (OFR) vs. Fractional flow reserve (FFR) was evaluated on Correlation between OFR and FFR (r=0.84, p=<0.001). Optical flow ratio (OFR) was highly correlated with fractional flow reserve (FFR) (r=0.84, P<0.001) and showed good agreement in stent-treated arteries immediately after percutaneous coronary intervention.