Key result
Omron HEM-711 DLX passes the European Society of Hypertension validation protocol for clinical use.
Why the study?
The accuracy of the Omron HEM-711 DLX home blood pressure monitor needed validation according to the European Society of Hypertension International Protocol.
Does the Omron HEM-711 DLX home blood pressure monitor accurately measure blood pressure compared to the European Society of Hypertension International Protocol standard in patients?
Does the Omron HEM-711 DLX home blood pressure monitor accurately measure blood pressure compared to the European Society of Hypertension International Protocol standard in patients?
The Omron HEM-711 DLX home blood pressure monitor meets the European Society of Hypertension International Protocol standards and can be recommended for patient use, though individual validation is advised.
Answers are grounded in this paper and cite their sources.
Device meets ESH protocol criteria; supports cautious home use but requires individual patient validation before broader adoption.
The Omron HEM-711 DLX home Blood pressure monitor was tested using the European Society of Hypertension International Protocol in 33 patients. The average error was 0.9+/-5.2 (SD) (range: -12 to 10 mmHg)/-0.8+/-5.6 (-19 to 8). It can be recommended for use by patients. With all automatic devices it is, however, important to validate its accuracy in each individual patient who uses it.
No takes yet. Share an insight, caveat, or question.
Grim et al. (2008) studied this question. The Omron HEM-711 DLX home blood pressure monitor passed the European Society of Hypertension protocol with an average error of 0.9+/-5.2 and -0.8+/-5.6 mmHg and is recommended for patient use.
Synapse has enriched 10 related papers. Consider them for comparative context: