Answers are grounded in this paper and cite their sources.
A fully automated analyzer for methylated L-arginine metabolites [N,N-dimethyl-L-arginine (ADMA), N-methylarginine (NMMA) and N,N'-dimethyl-L-arginine (SDMA)] by high-performance liquid chromatography with post-column fluorescence derivatization was developed. This system consists of an on-line extraction, a separation on a reversed phase ion-pair chromatograph, a post-column derivatization by o-phthaladehyde (OPA) and thiol reaction, and fluorescence detection. NMMA, ADMA and SDMA were separated in 40 min with isocratic elution by a combination of octanoate and cyclohexane carboxylate as ion-pair reagents. The eluate was monitored at 450 nm with excitation at 337 nm. The calibration curves for NMMA, ADMA and SDMA showed linearity over the range from 0.05 micromol l(-1) (0.5 pmol on column) to 5.0 micromol l(-1) (50 pmol on column). This method does not require any time-consuming pre-treatment and requires only 10 microl of plasma sample for assay.
No takes yet. Share an insight, caveat, or question.
Dobashi et al. (2001) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: