Key result
ECG-gating with pulse progressive read-out improves experimental angiocardiographic image quality.
Why the study?
The effects of short pulse width acquisition and ECG-gating on image quality in digital angiocardiography were not well established.
Does ECG-gating and short pulse width acquisition improve image quality in digital angiocardiography in an animal model?
Does ECG-gating and short pulse width acquisition improve image quality in digital angiocardiography in an animal model?
ECG-gating combined with short pulse width acquisition (1-5 ms) and pulse progressive read-out significantly improves image quality in digital angiocardiography.
Answers are grounded in this paper and cite their sources.
May guide human protocol development; leaves open clinical translation of ECG-gated angiocardiography.
Bogren HG, Seibert JA, Hines HH, Porter BA. The beneficial effects of short pulse width acquisition and ECG-gating in digital angiocardiography. Invest Radiol 1984;19:284-290. Experimental intravenous left ventriculography and coronary angiography after aortic root injection were performed in 13 animals. Various image acquisition and digital image processing techniques were employed. These included single mask mode subtraction, integrated mask mode subtraction, ECG-gated image acquisition, TV camera read-out in interlaced mode, and pulse progressive read-out with 1- to 5-ms pulse width. High quality images were obtained consistently only after ECG-gating and use of 1- to 5-ms pulse width with pulse progressive read-out of the TV camera. Integration of four to eight background as well as contrast images all from the same phase of the cardiac cycle further improved image quality.
No takes yet. Share an insight, caveat, or question.
Bogren et al. (1984) studied Experimental angiocardiography (n=13). ECG-gated image acquisition with 1- to 5-ms pulse width vs. Other image acquisition techniques was evaluated on Image quality. ECG-gating combined with 1- to 5-ms pulse width and pulse progressive read-out consistently produced high-quality images in experimental angiocardiography.
Synapse has enriched 10 related papers. Consider them for comparative context: