Key result
PS3 system shows 1-year biocompatibility and sustained septal-lateral dimension reductions in preclinical testing.
Why the study?
The long-term durability and safety of emerging percutaneous devices for functional mitral regurgitation remain largely unknown.
Does the PS3 system demonstrate long-term safety and durability in a healthy ovine model?
Does the PS3 system demonstrate long-term safety and durability in a healthy ovine model?
Answers are grounded in this paper and cite their sources.
The PS3 system demonstrated excellent biocompatibility, safety, and sustained reduction in mitral annular dimension up to 1 year in a healthy ovine model.
OBJECTIVES: Chronic implants of the PS(3) system were conducted in an ovine model to assess durability and safety at up to 1 year follow-up. BACKGROUND: The long-term durability and safety of emerging percutaneous devices for functional mitral regurgitation remain largely unknown. METHODS: The PS(3) system (consisting of interatrial septal and great cardiac vein devices connected by an adjustable suture bridge) was placed in eight healthy adult sheep. The mitral annular septal-lateral dimension in systole (SLS) was acutely reduced by 15-20%. Animals were sacrificed at up to 12 months postimplant and characterized by intracardiac echocardiography, cardiac computed tomography (CT), and histopathology. In vivo forces exerted on the PS(3) bridge were measured by means of a novel load cell catheter. RESULTS: At 3, 6, and 12 months after implantation, intracardiac echocardiographic and CT showed the PS(3) systems to be intact without erosion and with overall sustained reductions in the SLS. Histopathologic assessment revealed each component correctly deployed in its respective target site without evidence of erosion, thrombus, or device fracture. The SLS was 26.5 +/- 1.7 mm preimplant, 22.0 +/- 1.4 mm post-PS(3) (17.0% reduction), and 22.0 +/- 2.1 mm at latest follow-up. Mean forces exerted on the bridge in vivo ranged from 1.16 N to 1.87 N. CONCLUSIONS: The PS(3) System demonstrated excellent biocompatibility without evidence of erosion, thrombosis, or perforation at up to one-year follow-up in this chronic healthy ovine model. Forces exerted in the PS(3) system were relatively modest and should contribute to the durability of the device.
No takes yet. Share an insight, caveat, or question.
Rogers et al. (2008) studied Healthy ovine model for functional mitral regurgitation (n=8). PS3 system (percutaneous septal sinus shortening) was evaluated on Durability and safety (intactness, erosion, thrombus, device fracture, and septal-lateral dimension reduction). The PS3 System demonstrated excellent biocompatibility without erosion, thrombosis, or perforation at up to 1 year in a healthy ovine model, with sustained reductions in septal-lateral dimension.
Synapse has enriched 10 related papers. Consider them for comparative context: