Key result
Subcutaneous transvenous coil leads successfully lower defibrillation thresholds in complex ICD cases.
Why the study?
In complex cases of implantable cardioverter defibrillator implantation, additional subcutaneous array leads are commonly used to lower the defibrillation threshold.
Does the use of transvenous coil electrodes as single element subcutaneous array leads lower the defibrillation threshold in complex ICD cases?
Case Report
Does the use of transvenous coil electrodes as single element subcutaneous array leads lower the defibrillation threshold in complex ICD cases?
Answers are grounded in this paper and cite their sources.
Transvenous coil electrodes can be successfully repurposed as single element subcutaneous array leads to lower defibrillation thresholds in complex ICD cases.
For implantable cardioverter defibrillator a 10 J safety margin between the defibrillation threshold (DFT) and the maximum output of the device is intended. In complex cases, the additional placement of a subcutaneous array lead is a common strategy for lowering the DFT. We report the successful use of transvenous coil electrodes as single element subcutaneous array leads in order to lower the DFT.
No takes yet. Share an insight, caveat, or question.
Juchem et al. (2009) conducted a case report in High defibrillation threshold in implantable cardioverter defibrillator patients. Transvenous coil electrodes as single element subcutaneous array leads was evaluated on Lowering the defibrillation threshold (DFT). Transvenous coil electrodes were successfully used as single element subcutaneous array leads to lower the defibrillation threshold in complex implantable cardioverter defibrillator cases.
Synapse has enriched 10 related papers. Consider them for comparative context: