Key result
Prototype TAVR prosthesis matches clinical bioprosthesis hydrodynamic performance with comparable transvalvular gradients.
Why the study?
The study was motivated by the need to evaluate the hydrodynamic function of a minimally invasive transcatheter aortic valve prosthesis prototype under physiological flow conditions.
Supports prototype development; leaves open in vivo performance and clinical translation.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Ovcharenko et al. (2016) studied Aortic valve prosthesis evaluation. Prototype transcatheter aortic valve prosthesis vs. 3F Enable bioprosthesis was evaluated on Transprosthesis pressure gradient (mmHg) (p=0.37). The prototype transcatheter aortic valve prosthesis demonstrated hydrodynamic performance comparable to a clinical seamless bioprosthesis, with a mean transprosthesis pressure gradient of 18.01 mmHg versus 19.72 mmHg (p=0.37).